The injection resolution is dependent on the volume of the syringe that is installed, as outlined in the table on the previous page. For example, with a 25µl syringe installed, one complete rotation of the micrometer (25 divisions) yields a displacement equivalent to a volume of 267nl and turning the micrometer one division (0.001in) yields 10.6nl.
(1) Please specify the syringe volume that you will be using.
SyringeVolume Volume Per Revolution Volume Per Division 10µl 106nl 4.2nl 25µl 267nl 10.6nl 50µl 529nl 21.2nl 100µl 1.06µl 42.3nl 250µl 2.65µl 105.8nl 500µl 5.29µl 211.7nl 1000µl 10.58µl 423.4nl
Dimensions 12in x 3.75in x 4in | 30cm x 9.5cm x 10cm
Weight 2.5 lbs | 1 kg
|